BlueKit series cell residual human IL-2 ELISA detection kits use the double-antibody sandwich method to detect IL-2 protein in samples.
| Specification | 96 Test |
Assay range | 7.81-500 pg/mL |
Detection sensitivity | 0.59 pg/mL |
Precision | CV% ≤ 10%, RE% ≤ ±15% |
| Validity period | 12 months |
| Storage conditions | 4℃ |
| LOQ | 7.81pg/mL |
| LOD | 0.13 pg/mL |
| Concentration of standard (pg/mL) | OD value1 | OD value2 | Mean value |
| 0.00 | 0.0214 | 0.0215 | 0.0215 |
| 7.81 | 0.0674 | 0.0665 | 0.0670 |
| 15.63 | 0.1140 | 0.1154 | 0.1147 |
| 31.25 | 0.2154 | 0.2062 | 0.2108 |
| 62.50 | 0.4035 | 0.3916 | 0.3976 |
| 125.00 | 0.7904 | 0.7754 | 0.7829 |
| 250.00 | 1.5240 | 1.4780 | 1.5010 |
| 500.00 | 2.6660 | 2.5960 | 2.6310 |
Standard curve: