This product utilizes a double-antibody sandwich method to detect residual Pichia pastoris host cell proteins (HCP) in samples. A microtiter plate pre-coated with capture antibody specific to Pichia pastoris HCP is used. After adding samples and standards, the capture antibody specifically binds to the Pichia pastoris HCP. Following a wash step to remove impurities, an enzyme-labeled antibody is added. After another wash step to remove unbound enzyme-labeled antibodies, a substrate is added for color development. The reaction is then terminated, and the absorbance is read. The absorbance value of the sample is directly proportional to the amount of Pichia pastoris HCP detected in the sample. By comparing the absorbance to a standard curve and multiplying by the corresponding dilution factor, the residual amount of Pichia pastoris HCP in the sample can be determined.
High Sensitivity and Accurate Quantification
The detection range of 1.5625–100 ng/mL ensures precise measurement of residual Pichia pastoris host cell proteins (HCPs) using a validated double-antibody sandwich ELISA method, meeting stringent quality control requirements for biologics manufacturing.
Regulatory Compliance and Standardized Workflow
Compliant with Chinese Pharmacopoeia (2020 Edition) and ICH Q6B guidelines, the kit includes pre-coated antibody plates and calibrated standards, with a standardized protocol (e.g., 4PL curve fitting) to ensure reproducibility and regulatory acceptance.
Efficient Operation and Reliable Stability
The total assay time is approximately 2 hours and 50 minutes, with opened reagents stable for 1 month (2–8°C). Detailed troubleshooting guidance (e.g., for washing or color development issues) ensures robust performance in routine lab testing.
1.Preparation of Reagents:
Take the reagent kit out of the refrigerator and allow it to equilibrate to room temperature for at least 30 minutes, ensuring all required reagents for the experiment are ready. Mark the opening date on the reagent kit and prioritize using the oldest opened kit. Kits from the same batch number can be used together, but kits from different batch numbers should not be mixed. Return any unused strip plates back to their sealed bag and store at 2–8°C for future use.
Prepare the Wash Solution (1×): Dilute an appropriate amount of 20× Wash Buffer with deionized water to achieve a 1× solution. If crystals are present in the 20× Wash Buffer, place it at room temperature or in a 37°C water bath, gently shaking until the crystals are fully dissolved before dilution.
2.Standard Curve Preparation:
The Pichia HCP calibration standard does not require dilution and can be used directly.
Sample Solution Preparation:
Dilute the sample using Sample Diluent Buffer to bring it within the range of the standard curve, and mix thoroughly.
Spiked QC Solution Preparation:
Take 120 µL of the sample solution and add 120 µL of the 25 ng/mL standard, mix thoroughly.
Washing the Plate:
Wash the plate with 1× Wash Solution, adding 300 µL per well. Dry any residual liquid from the wells by inverting and tapping the plate gently. Repeat this process three times.
Sample Incubation:
Add 100 µL of the standard, sample, or spiked QC to the corresponding wells. Seal the plate with a sealing film and incubate at 37°C with shaking at 300 rpm for 1 hour. (Note: Ensure the plate is properly sealed during incubation; incomplete sealing or lack of sealing can lead to evaporation, causing errors in the results.)
3.Washing the Plate Again:
After incubation, let the plate sit at room temperature for 3-5 minutes. Carefully remove the sealing film and discard the liquid in the wells. Wash the plate with 1× Wash Solution, adding 300 µL per well. Dry any residual liquid by inverting and tapping the plate. Repeat this process three times.
(Note: If washing manually, dispense the wash buffer above the wells without touching the well walls with the pipette tips. After each wash, let the plate sit for 30 seconds and gently shake before inverting the plate to dry the wells. Use new absorbent paper or a clean section of the same paper each time.)
Preparation of HRP-conjugated Antibody:
Dilute the 100× Anti-Pichia HCP-HRP antibody using Antibody Dilution Buffer to achieve a 1× working solution. For example, add 100 µL of 100× Anti-Pichia HCP-HRP to 9.9 mL of Antibody Dilution Buffer.
4.Antibody Incubation:
Add 100 µL of the diluted HRP-conjugated antibody to each well, seal the plate, and incubate at 37°C with shaking at 300 rpm for 1 hour. Take Color Reagent A and B out of the 4°C refrigerator and allow them to equilibrate to room temperature.
Washing the Plate Again:
After incubation, let the plate sit at room temperature for 3-5 minutes. Carefully remove the sealing film and discard the liquid. Wash the plate five times with 1× Wash Solution, adding 300 µL per well each time. Dry any residual liquid from the wells.
5.Color Development:
Mix Color Reagent A and B in a 1:1 ratio and add 100 µL of the mixture to each well. Seal the plate and incubate at room temperature, protected from light, for 20 minutes.
6.Termination:
Add 50 µL of Stop Solution to each well to terminate the reaction. After stopping the reaction, read the plate as soon as possible. (Note: It is recommended to set the microplate reader to perform a 5-10 second shaking before reading.)
Reading the Plate:
Within 20 minutes, measure the absorbance at 450 nm with a reference wavelength at 630 nm using a microplate reader.
| Serial No. | Problem Description | Possible Causes | Corresponding Countermeasures |
| 1 | Gradient difference of standard curve | Inaccurate pipetting or liquid adding | Check pipettes and tips |
| Incomplete washing of microplate | Ensure the number of washing cycles and the volume of washing solution per well | ||
| 2 | No color development or very weak color development in both standard curve and samples | Too short incubation time | Ensure sufficient incubation time |
| Incorrect experimental temperature | Use the recommended incubation temperature | ||
| Omission of a component, especially detection antibody or enzyme | Check experimental records and remaining reagents. Verify labels before each liquid addition | ||
| Expired reagents | Use products within the validity period | ||
| Inactivation or loss of standards/antibodies/enzymes/chromogenic substrates | Store correctly and replace with new standards/antibodies/enzymes/chromogenic substrates | ||
| Failure to add the next reaction solution in time after washing and pat - drying the microplate | Add the next reaction solution immediately after washing and pat - drying the plate | ||
| 3 | No color development or very weak color development in standard curve, but color development in samples | Insufficient or no vortexing during serial dilution of standards | Use vortex mixing during dissolution and dilution |
| Sample coloration | |||
| 4 | Low OD value reading | Incorrect setting of microplate reader | Check the wavelength and filter settings on the microplate reader |
| Turn on the microplate reader in advance for preheating before reading | |||
| Improper washing operation of microplate: e.g., too many washing cycles, too long residence time after adding washing solution, etc. | Wash the plate according to the method recommended in the instruction manual | ||
| 5 | Large coefficient of variation | Mismatch between pipette and tips | Replace pipette tips |
| Poor accuracy of pipette | Regularly calibrate and test the pipetting equipment | ||
| Check the bottom of the microplate | Check if there is residual liquid and fingerprints on the bottom of the microplate | ||
| Inconsistent pipetting operation | Practice pipetting repeatedly to maintain consistency in pipetting operation | ||
| Abnormality inside the plate wells | Confirm no foreign matter in the plate wells before sample addition and no air bubbles after sample addition | ||
| 6 | High background value | Incomplete washing of microplate | Wash the plate according to the method recommended in the instruction manual |
| If using an automatic plate washer, check if all liquid addition ports and waste liquid discharge ports are blocked | |||
| If washing the plate by hand, appropriately increase the number of washing cycles | |||
| Contamination of shared reagents: e.g., ultrapure water | Replace with non - contaminated reagents and re - prepare | ||
| Contamination of shared instruments: e.g., pipettes, centrifuges | Use dedicated pipettes and sterile filter tips | ||
| Unclean operating environment, mixed use of ELISA test operation area with cell culture and disruption areas | Separate the test operation areas | ||
| Incorrect reagent matching: e.g., incorrect dilution ratio of washing solution or detection antibody | Re - prepare according to the correct dilution ratio | ||
| Too long reaction time | Immediately use the stop solution to terminate the reaction when the color development of the microplate is sufficient for absorbance reading, and appropriately shorten the color development time if necessary | ||
| Color development reaction not protected from light | Conduct the color development reaction under light - proof conditions | ||
| 7 | Experimental results deviate greatly from the reference performance parameters | Improper storage of kit | Store relevant reagents as required in the instruction manual |
| Expired reagents | Confirm that the kit and its components are within the validity period | ||
| Failure to strictly follow the instruction manual during the experiment | Conduct practical operation training for experimenters before the experiment to ensure the experiment proceeds smoothly | ||
| Strictly control key experimental steps, such as use concentration, sample volume, incubation time, etc., and do not replace the instruction manual with empirical judgment | |||
| 8 | Positive result in negative control | Contamination of samples or reagents, or improper operation during sample addition leading to cross - contamination due to splashing between adjacent wells | Replace reagents and operate carefully |
| Incomplete washing of microplate | Pour out the antibody solution before washing the plate, then fill the plate wells with washing solution to ensure sufficient washing | ||
| 9 | Large difference in calculated sample values when samples are diluted at different gradients | Strong matrix effect of samples | Select the dilution factors where the calculated sample values are close under two dilution gradients (for samples with relatively high content of target protein) |
Purpose: Quantitative measurement of Pichia pastoris host cell protein (HCP) residuals in biologics intermediates and final products.
Feature: Double-antibody sandwich ELISA with high specificity (1.5625–100 ng/mL detection range).
Technical Effort: Validated 4PL curve fitting (CV <5% for replicates).
Rigorous QC: Spike recovery (95–107%), inter-assay precision (CV <10%).
Application: Lot release testing for recombinant protein therapeutics.
Purpose: Ultra-sensitive detection of E. coli residual DNA in biologics (e.g., vaccines, gene therapies).
Feature: qPCR with TaqMan probe technology (fg/μL-level sensitivity).
Technical Effort: Standard curve validation (R²>0.99, 90–110% efficiency).
Application: Compliance testing per USP/EP/ChP guidelines.
Purpose: Dynamic monitoring of HCP levels during upstream/downstream bioprocessing.
Feature: Ready-to-use ELISA kits for multiple host systems (e.g., 293T, Pichia).
Technical Effort: Pre-coated plates with batch-to-batch consistency.
Application: In-process control (IPC) for cell culture and purification steps.
Application Scenario
A biologics manufacturer required precise quantification of Pichia pastoris HCP residuals in a recombinant protein drug to comply with EMA and FDA regulations.
Technical Challenge
Detection of low-abundance HCPs (1.56–100 ng/mL) in complex samples with high product concentrations, while minimizing matrix interference.
Solution Implemented
• Used pre-coated 96-well plates (HG-HCP005) with optimized diluent buffers.
• Validated 4PL curve fitting (R²>0.99) on a SpectraMax microplate reader.
• Included spike recovery controls (25 ng/mL) and inter-plate QC samples.
Outcome & Validation Data
• Achieved LOD of 0.8 ng/mL (CV <5%).
• All batches met ≤5 ng/mg HCP specification (ICH Q6B).
• Inter-assay precision: CV 6.8% (n=20).
• Spike recovery: 95–107% across 3 production sites.
Application Scenario
A CAR-T cell therapy developer needed to monitor HEK293 residual DNA during viral vector production.
Technical Challenge
Quantifying fragmented DNA (<200 bp) in lentiviral supernatants with high sensitivity (fg/μL).
Solution Implemented
• Deployed qPCR kit (HG-HD003) with TaqMan probes targeting SV40 promoter.
• Automated extraction on QIAcube to reduce operator variability.
• Included DNA fragmentation analysis (HG-HF002) for size distribution.
Outcome & Validation Data
• Detected 50 fg/μL DNA (S/N >3.5).
• Reduced false positives via fragmentation profiling (95% fragments <500 bp).
• Inter-operator CV: 3.9% (n=12).
• Compliance with USP <1043> for cell-based therapies.
| Product Code Valid for 12 months | Product Specifications | Expiration Date: |
| HG-HCP005 | 96T | 12 months |
| HG-HCP005-S | 48T |