Medical Consumables and Lab Consumables OEM Manufacturer
Medical Consumables and Lab Consumables OEM Manufacturer

RCL(VSVG) Gene Copy Number Detection Kit (qPCR)

This kit is designed for the quantitative detection of RCL gene copy number in the genome of CAR-T cells prepared by using HIV-1 lentiviral vector technology.


This kit adopts the fluorescent probe method and multiplex PCR method to detect the DNA sequence related to integration or expression function on the transfer plasmid, and the VSVG gene copy number in the sample can be calculated. The kit is a rapid, specific and reliable device.


Contact
Get A Quote
HG-RC001
2

Specification of RCL(VSVG) Gene Copy Number Detection Kit (qPCR)


Assay range

1.00×101~1.00×105copies/μL

Limit of quantitation

8 copies/μL

Limit of detection

1 copies/μL

Precision

CV%≤15%



Data of RCL(VSVG) Gene Copy Number Detection Kit (qPCR)

Concentration
(copies/μl)
Log10

Concentration
Ct Value(1)Ct Value(2)Ct Value(3)Ct Mean ValueCV
1.00E+055.0020.2720.16        20.13         20.18         0.36%
1.00E+044.0023.3523.2123.2723.280.32%
1.00E+033.0026.9326.8526.9626.910.21%
1.00E+022.0030.3130.3130.2030.270.21%
1.00E+011.0033.5433.333.1933.340.53%
Amplification efficiency99.60%


Standard curve:

Data of RCL(VSVG) Gene Copy Number Detection Kit (qPCR)

Products
More
Bring your cellular therapy products to the next milestone sooner
Contact Us
Contact
86 198 5711 6962 info@hillgene.com
Avenue Building 4, Yuewang Wisdom Valley, 1463 Wuzhong Avenue, Wuzhong District, Suzhou China
Ask Us Anything!