BlueKit series human IFN-γ ELISA detection kits use the double-antibody sandwich method to detect IFN-γ protein in samples.
| Specification | 96 Test |
Assay range | 4.69-300pg/mL |
Detection sensitivity | 0.36pg/mL |
Precision | CV% ≤ 10%, RE% ≤ ±15% |
| Validity period | 12 months |
| Storage conditions | 4℃ |
| LOQ | 1.37pg/mL |
| LOD | 0.36pg/mL |
| Concentration of standard (pg/mL) | OD value1 | OD value2 | Mean value |
| 300.00 | 4.0655 | 4.0291 | 4.0473 |
| 150.00 | 3.1760 | 3.2300 | 3.2030 |
| 75.00 | 2.1280 | 2.1770 | 2.1525 |
| 37.50 | 1.2630 | 1.2960 | 1.2795 |
| 18.75 | 0.6812 | 0.7103 | 0.6958 |
| 9.38 | 0.3773 | 0.3778 | 0.3776 |
| 4.69 | 0.1919 | 0.2089 | 0.2004 |
| 0.00 | 0.0218 | 0.0220 | 0.0219 |
Standard curve: