BlueKit series cell residual human IL-15 ELISA detection kits use the double-antibody sandwich method to detect IL-15 protein in samples.
| Specification | 96 Test |
Assay range | 3.91-250pg/mL |
Detection sensitivity | 0.24pg/mL |
Precision | CV% ≤ 10%, RE% ≤ ±15% |
| Validity period | 12 months |
| Storage conditions | 4℃ |
| LOQ | 0.69 pg/mL |
| LOD | 0.14pg/mL |
| Concentration of standard (pg/mL) | OD value1 | OD value2 | Mean value |
| 0.00 | 0.0537 | 0.0513 | 0.0525 |
| 3.91 | 0.1116 | 0.1070 | 0.1093 |
| 7.81 | 0.1822 | 0.1820 | 0.1821 |
| 15.63 | 0.3069 | 0.3381 | 0.3225 |
| 31.25 | 0.6350 | 0.6030 | 0.6190 |
| 62.50 | 1.2330 | 1.0960 | 1.1645 |
| 125.00 | 2.1470 | 2.0170 | 2.0820 |
| 250.00 | 3.1130 | 3.0550 | 3.0840 |
Standard curve: