Specification of 293T Cell Residual DNA Detection Kit (qPCR)


Specification100 Reactions

Assay range

3.00×10¹~3.00×10⁵ fg/μL

Limit of quantitation

3.00×10¹ fg/μL

Precision

CV%≤15%

Validity period18 months
Storage conditions-20℃
LOQ30 fg/μL
LOD3 fg/μL

Datasheet of 293T Cell Residual DNA Detection Kit (qPCR)

Concentration
(fg/μl)
Log
Concentration
Ct Value(1)Ct Value(2)Ct Value(3)Ct Mean ValueRecovery rate
3.00E+055.4823.6423.7123.5923.6592.00%
3.00E+044.4826.7926.8526.7826.8197.00%
3.00E+033.4830.0830.1130.0330.07101.00%
3.00E+022.4833.8933.7333.7633.7999.00%
3.00E+011.4836.0336.5236.3936.3191.00%
Amplification efficiency97.90%


Standard curve:

Datasheet of 293T Cell Residual DNA Detection Kit (qPCR)

Products
Bring your cellular therapy products to the next milestone sooner
Contact Us
Contact
1-805-304-6109 info@hillgene.com
Avenue de l'Hôpital 11, 4000 Liège (Sart-Tilman)
Ask Us Anything!
On June 26, Professor Raju Kucherlapati, a world-renowned expert in genetics and medicine and member of the U.S. National Academy of Medicine (NAM), visited UPMED Biotech with his delegation.
Hillgene News & Blog