The BaEV Gene Copy Number Detection Kit is a specialized kit for quantitative detection of BaEV gene copy number.
This kit quantitatively detects the copy number of BaEV gene in the sample based on the fluorescence probe method. This kit is rapid, specific and reliable in performance.
Assay range | 2.00×101~2.00×106copies/μL |
Limit of quantitation | 20 copies/μL |
Limit of detection | 2 copies/μL |
Precision | CV%≤15% |
Concentration (copies/μL) | Log10 Concentration | Ct Value(1) | Ct Value(2) | Ct Value(3) | Ct Mean Value | CV |
2.00E+05 | 5.30 | 20.86 | 21.01 | 20.74 | 20.87 | 0.64% |
2.00E+04 | 4.30 | 24.05 | 24.17 | 23.83 | 24.02 | 0.71% |
2.00E+03 | 3.30 | 27.81 | 27.50 | 27.56 | 27.62 | 0.60% |
2.00E+02 | 2.30 | 30.95 | 30.80 | 30.73 | 30.83 | 0.36% |
2.00E+01 | 1.30 | 34.24 | 34.42 | 33.79 | 34.15 | 0.96% |
2.00E+00 | 0.30 | 37.41 | 37.81 | 37.72 | 37.64 | 0.55% |
Amplification efficiency | 99.60% |
Standard curve: